產(chǎn)品列表PRODUCTS LIST
甲基環(huán)己烷C6H11CH3氣體檢測管113SB METHYL CYCLOHEXANE 100-1600ppm,氣體檢測管的原理及使用,操作簡(jiǎn)便。使用時(shí)僅有采樣和測試結果顯示兩步,而這兩步又近乎是同時(shí)進(jìn)行的,為專(zhuān)業(yè)檢驗人員提供了極大方便,只要操作人員按照使用說(shuō)明書(shū)操作方法進(jìn)行測試就可以應用。
更新時(shí)間:2024-07-14
在線(xiàn)留言甲基環(huán)己烷C6H11CH3氣體檢測管113SB METHYL CYCLOHEXANE 100-1600ppm
Tube 13SB
1. PERFORMANCE
1) Measuring range :50-1,400 ppm
Number of pump strokes 1(100ml)
2) Sampling time :1.5 minutes/1 pump stroke
3) Detectable limit :5 ppm
4) Shelf life :2 years
5) Operating temperature :0~40℃
6) Temperature compensation :Necessary(See "TEMPERATURE CORRECTION TABLE")
7) Reading :Direct reading from the scale calibrated by 1 pump stroke
8) Colour change :Orange→Yellowish green
2. RELATIVE STANDARD DEVIATION
RSD-low:10% RSD-mid.:5% RSD-high:5%
3. CHEMICAL REACTION
Chromium oxide is reduced.
CH3(CH2)4CH3+Cr6++H2SO4→Cr3+
4. CALIBRATION OF THE TUBE
GAS CHROMATOGRAPHY
5. INTERFERENCE AND CROSS SENSITIVITY
GAS To pump
1400
1200
1000
800
600
400
200
100
50
ppm
????
????
????
????
????
????
113SB n-HEXANE
Substance Interference Coexistence
Aromatic hydrocarbons FIG.1,2 Black stain is produced. Higher readings are given.
Paraffin (morefi nth haynd Cro3c)arbons Similar stain is produced. 〃
In presence of Alcohols, Ketones or Esters less than 6%, the accuracy of readings is not affected.
0
500
Benzene〔ppm〕?
FIG.1 Influence of Benzene
n-Hexane indicating value 〔ppm〕
50 100 0
500
Toluene〔ppm〕?
FIG.2 Influence of Toluene
n-Hexane indicating value 〔ppm〕 100 150 50
TEMPERATURE CORRECTION TABLE
Scale True Concentration (ppm)
Readings 0℃ 10℃ 20℃ 30℃ 40℃
(ppm) (32°F) (50°F) (68°F) (86°F) (104°F)
1,400 1,730 1,530 1,400 1,270 1,180
1,200 1,470 1,320 1,200 1,090 1,010
1,000 1,230 1,100 1,000 910 840
800 980 870 800 720 670
600 740 660 600 550 500
400 470 430 400 360 330
200 220 210 200 180 170
100 100 100 100 100 100
上一篇:190U醋酸甲氧乙酯氣體檢測管Methyl cellosolve acetate 3-120ppm
下一篇:199U甲基環(huán)己醇CH3C6H10OH氣體檢測管199U METHYL CYCLOHEXANOL 5-200
返回列表>>相關(guān)產(chǎn)品