1,3-二氯丙烷氣體檢測管194S 1,3-DICHLOROPROPANE 10-500ppm,總代理銷(xiāo)售日本光明理化學(xué)工業(yè)株式會(huì )社北川式氣體檢測管,提供日本光明理化學(xué)工業(yè)株式會(huì )社北川式氣體檢測管介紹,日本光明理化學(xué)工業(yè)株式會(huì )社北川式氣體檢測管應用,日本光明理化學(xué)工業(yè)株式會(huì )社北川式氣體檢測管價(jià)格,日本光明理化學(xué)工業(yè)株式會(huì )社北川式氣體檢測管說(shuō)明書(shū),日本光明理化學(xué)工業(yè)株式會(huì )社北川式氣體檢測管售后服務(wù),日本光明理化學(xué)工業(yè)株式會(huì )社北川式氣體檢測管中國總代理
更新時(shí)間:2024-07-14
在線(xiàn)留言1,3-二氯丙烷氣體檢測管194S 1,3-DICHLOROPROPANE 10-500ppm
Tube No.
194S
1. PERFORMANCE
1) Measuring range :10-500 ppm
Number of pump strokes 1(100ml)
2) Sampling time :1.5 minutes/1 pump stroke
3) Detectable limit :1 ppm
4) Shelf life :1 year (Necessary to store in a refrigerated place ; 0~10℃)
5) Operating temperature :0~40℃
6) Temperature compensation :Necessary(See "TEMPERATURE CORRECTION TABLE")
7) Reading :Direct reading from the scale calibrated by 1 pump stroke
8) Colour change :White→Orange
2. RELATIVE STANDARD DEVIATION
RSD-low:15% RSD-mid.:15% RSD-high:10%
3. CHEMICAL REACTION
By decomposing with an Oxidizer, chlorine is produced. Yellowish Orthoquinone is produced by reacting
between this Chlorine and o-Toulidine.
CH2CICH2CH2CI+I2O4+CrO3+H2SO4→CI2
4. CALIBRATION OF THE TUBE
GAS CHROMATOGRAPHY
5. INTERFERENCE AND CROSS SENSITIVITY
??????????
GAS To pump
DETECTOR TUBE
(White)
PRETREAT TUBE
500
400
300
200
100
50
30
10
ppm
100ml
194S 1,3-DICHLOROPROPANE
Substance Interference Coexistence
Alcohols The accuracy of readings is not affected. The accuracy of readings is not
affected.
Esters 〃〃
Aromatic hydrocarbons 〃〃
Aliphatic hydrocarbonFsIG.1 〃〃
Halogenated hydrocarFbIoGn.s2 Similar stain is produced. Higher readings are given.
0
300
400
500
200
100
Trichloroethylene〔ppm〕?
FIG. 2 Influence of Trichloroethylene
上一篇:132SC1,3-二氯丙烷氣體檢測管132SC 1,3-Dichloropropane 1-10ppm
下一篇:190U四氫萘C10H12氣體檢測管190U Dicyclopentadiene 2-60ppm