產(chǎn)品列表PRODUCTS LIST
157SB METHYL BROMIDE (-) 157SB甲基溴,CH3CHCLCH2Cl又稱(chēng)氯化丙烯,1,2-二氯丙烷;氯化丙烯;DCP;1,2-Dichloropropane,20-250ppm 總代理銷(xiāo)售日本光明理化學(xué)工業(yè)株式會(huì )社北川式氣體檢測管,提供日本光明理化學(xué)工業(yè)株式會(huì )社北川式氣體檢測管介紹,日本光明理化學(xué)工業(yè)株式會(huì )社北川式氣體檢測管應用,日本光明理化學(xué)工業(yè)株式會(huì )社北川式氣體檢測管價(jià)格,日本光明理化學(xué)工業(yè)株式會(huì )社北川式氣體檢測管說(shuō)明書(shū),日本光明理化學(xué)工業(yè)株式會(huì )社北川式氣體檢測管售后www.yd17.cn
更新時(shí)間:2024-07-14
在線(xiàn)留言甲基溴氣體檢測管157SB METHYL BROMIDE 20-250ppm
157SB
1. PERFORMANCE
1) Measuring range :10-500ppm
Number of pump strokes 1(100ml)
2) Sampling time :2 minutes/1 pump stroke
3) Detectable limit :1 ppm
4) Shelf life :3 years(Necessary to store in a refrigerated place ; 0~10℃)
5) Operating temperature :0~40℃
6) Temperature compensation :Necessary(See "TEMPERATURE CORRECTION TABLE")
7) Reading :Direct reading from the scale calibrated by 1 pump stroke
8) Colour change :White→Reddish orange
2. RELATIVE STANDARD DEVIATION
RSD-low:15% RSD-mid.:10% RSD-high:10%
3. CHEMICAL REACTION
By decomposing with an Oxidizer, Bromine is produced. Bromine reacts with o-Toluidine and red Orthoquinone
is produced.
CH3Br+I2O5+CrO3+H2SO4→Br2
4. CALIBRATION OF THE TUBE
GAS CHROMATOGRAPHY
5. INTERFERENCE AND CROSS SENSITIVITY
10
30
50
100
150
200
300
400
500
ppm
100ml
GAS To pump
(Brown) (White)
PRETREAT TUBE DETECTOR TUBE
Tube No.
157SA METHYL BROMIDE
Substance ppm Interference ppm Coexistence
Chlorine 1 Similar stain is produced. 1 Higher readings are given.
Bromine 1 〃1 〃
Nitrogen dioxide 1 〃1 〃
Dichloromethane 500 〃500 〃
Chloroform 50 〃50 〃
Ethylene dibromide 50 〃50 〃
Trichloroethylene 50 〃50 〃
Tetrachloroethylene 50 〃50 〃
TEMPERATURE CORRECTION TABLE
Scale True Concentration (ppm)
Readings 0℃ 5℃ 10℃ 15℃ 20℃ 30℃ 40℃
(ppm) (32°F) (41°F) (50°F) (59°F) (68°F) (86°F) (104°F)
500 - - - 620 500 430 390
400 - - 630 475 400 360 325
300 - 675 430 345 300 280 260
200 580 320 235 210 200 190 175
150 260 185 160 150 150 140 130
100 120 110 100 100 100 95 90
50 50 50 50 50 50 50 50
185
1. PERFORMANCE
1) Measuring range :5-80 ppm 2.5-40 ppm 0.9-7.3 ppm
Number of pump strokes 1(100ml) 2(200ml) 4(400ml)
2) Sampling time :1.5 minutes/1 pump stroke
3) Detectable limit :1 ppm(200ml)
4) Shelf life :3 years(Necessary to store in a refrigerated place ; 0~10℃)
5) Operating temperature :0~40℃
6) Temperature compensation :Necessary(See "TEMPERATURE CORRECTION TABLE")
7) Reading :Direct reading from the scale calibrated by 1 pump stroke
8) Colour change :White→Yellow
2. RELATIVE STANDARD DEVIATION
RSD-low:15% RSD-mid.:10% RSD-high:10%
3. CHEMICAL REACTION
By decomposing with an Oxidizer, Bromine is produced. Bromine reacts with o-Toluidine and yellow
Orthoquinone is produced.
CH3Br+I2O5+CrO3+H2SO4→Br2
4. CALIBRATION OF THE TUBE
GAS CHROMATOGRAPHY
5. INTERFERENCE AND CROSS SENSITIVITY
5
10
15
20
30
40
60
80
ppm
100ml
GAS To pump
(Olive) (White)
PRETREAT TUBE DETECTOR TUBE
Tube No.
157SB METHYL BROMIDE
(NOTE)
When the concentration is below 5 ppm, 2 pump strokes can be used to determine the lower concentration.
Following formula is available for the actual concentration.
Actual concentration=1/2×Temperature corrected value
Substance Interference ppm Coexistence
Halogen Similar stain is produced. Higher readings are given.
Halogenated hydrocarbons 〃〃
Hexane FIG.1 200 Lower readings are given.
0
50
60
70
80
30
40
Hexane〔ppm〕?
FIG.1 Influence of Hexane
上一篇:180S二氯甲烷CH2Cl2氣體檢測管180S Dichloromethane 10-1000ppm
下一篇:132SC1,3-二氯丙烷氣體檢測管132SC 1,3-Dichloropropane 1-10ppm
返回列表>>相關(guān)產(chǎn)品