www.cgongxi.com-国产日产久久高清欧美一区WW,国产在线观看无码免费视频,国产日产久久高清欧美一区WW,国模欢欢炮交啪啪150P

產(chǎn)品列表PRODUCTS LIST

首頁(yè)>產(chǎn)品中心>日本光明北川式氣體檢測管>氣體快速檢測管>111U乙酸乙酯氣體檢測管111U ETHYL ACETATE 20-300ppm

乙酸乙酯氣體檢測管111U ETHYL ACETATE 20-300ppm

簡(jiǎn)要描述:

乙酸乙酯C6H10 111U ETHYL ACETATE (+) Cyclohexene 環(huán)己烯 20-300ppm,重要的是,不同的濃度意味著(zhù)被抽取的空氣中的氧氣的量不同,當有大量氧氣時(shí)(濃度較高)氧氣分子會(huì )前行的很遠(變白的區域很長(cháng))。然而當只有很少的氧氣時(shí)(濃度較低),就會(huì )很快穿過(guò)試劑與之反應而耗盡,這樣變白的區域就很短。

更新時(shí)間:2024-07-14

在線(xiàn)留言
乙酸乙酯氣體檢測管111U ETHYL ACETATE 20-300ppm

乙酸乙酯氣體檢測管111U ETHYL ACETATE 20-300ppm

Tube No.
111U
1. PERFORMANCE
1) Measuring range :20-1,000 ppm
Number of pump strokes 1(100ml)
2) Sampling time :1.5 minutes/1 pump stroke
3) Detectable limit :10 ppm
4) Shelf life : :2 years
5) Operating temperature :10~40℃
6) Temperature compensation :Necessary(See "TEMPERATURE CORRECTION TABLE")
7) Reading :Direct reading from the scale calibrated by 1 pump stroke
8) Colour change :Yellow→Brown
2. RELATIVE STANDARD DEVIATION
RSD-low:10% RSD-mid.:10% RSD-high:5%
3. CHEMICAL REACTION
Chromium oxide is reduced.
CH3CO2C2H5+Cr6++H2SO4→Cr3+
4. CALIBRATION OF THE TUBE
GAS CHROMATOGRAPHY
5. INTERFERENCE AND CROSS SENSITIVITY
GAS To pump
1000
800
600
400
200
100
50
20
??????????

????
?? ??
111U ETHYL ACETATE
TEMPERATURE CORRECTION TABLE
Scale True Concentration (ppm)
Readings 10℃ 15℃ 20℃ 25℃ 30℃ 35℃ 40℃
(ppm) (50°F) (59°F) (68°F) (77°F) (86°F) (95°F) (104°F)
1,000 1,330 1,200 1,000 830 700 600 500
800 1,080 950 800 670 550 470 400
600 830 720 600 500 420 380 300
400 570 480 400 330 280 240 200
200 280 240 200 170 140 120 100
100 150 120 100 80 70 60 50
50 70 60 50 40 30 20 10
20 30 25 20 15 10 7 4
Substance Interference Coexistence
Alcohols Similar stain is produced. Higher reading are given.
Esters 〃〃
Ketones 〃〃
Aromatic hydrocarbons 〃〃
Whole reagent is discoloured to It the top of brown stain is clear,
Paraffin hydrocarbons Pale brown. athffee catcecdu.racy of readings is not
Halogenated hydrocarbons FIG.1 〃〃
0
200
400
600
800
1000
Trichloroethylene〔ppm〕?
FIG.1 Influence of Trichloroethylene
Ethyl acetate indicating value 〔ppm〕
100 200 300
112
Tube No.
1. PERFORMANCE
1) Measuring range :5-60 ppm
Number of pump strokes 2(200ml)
2) Sampling time :3 minutes/2 pump strokes
3) Detectable limit :0.5 ppm
4) Shelf life : :2 years
5) Operating temperature :0~40℃
6) Temperature compensation :Necessary(See "TEMPERATURE CORRECTION TABLE")
7) Reading :Graduations printed on the tube are calibrated by Methyl acrylate at 2 pump strokes
and Ethyl acrylate concentration is determined by using a conversion chart.
8) Colour change :Yellow→Pale blue
2. RELATIVE STANDARD DEVIATION
RSD-low:15% RSD-mid.:10% RSD-high:10%
3. CHEMICAL REACTION
Chromium oxide is reduced.
CH2=CHCO2CH2CH3+Cr6++H2SO4→Cr3+
4. CALIBRATION OF THE TUBE
GAS CHROMATOGRAPHY
5. INTERFERENCE AND CROSS SENSITIVITY
GAS To pump
60
40
20
10
5
ppm
200ml
DEHUMIDIFIER
(White)
??????????

????
211U C ETHYL ACR